Performance Evaluation of Hematology Analyser Sysmex KX – 21 for Platelet and WBC Count

Performance Evaluation of Hematology Analyser Sysmex KX – 21 for Platelet and WBC Count

Authors

  • Muneeza Natiq
  • Rabia Ahmed
  • Noshin Wasim Yusuf

DOI:

https://doi.org/10.21649/akemu.v18i3.420

Abstract

Objective:  To evaluate the performance of Sysmex KX – 21 for the estimation of Platelet and WBC count in comparison with microscopy.

Study Design:  It is prospective cohort study.

Place and Duration:  It was conducted in the Patho-logy department, Allama Iqbal Medical College, from 7th Oct. 2009 to 6th Nov. 2009.

Materials and Methods:  Out of the total 1500 samp-les, 1458 (97%) samples after exclusions were finally evaluated for the study. On each sample, automated blood cell count was performed using Sysmex KX – 21 and the results were compared by peripheral smear examination for Platelet count, WBC count and its dif-ferential. The samples were from Medicine, Pediatrics and Oncology wards only.

Results:  Out of 1458 samples, WBC count of 1452 (99.5%) samples matched the total count on peripheral smear examination. 06 (0.5%) samples showed spur-ious WBC count due to the presence of nucleated

Downloads

How to Cite

Natiq, M., Ahmed, R., & Yusuf, N. W. (2013). Performance Evaluation of Hematology Analyser Sysmex KX – 21 for Platelet and WBC Count. Annals of King Edward Medical University, 18(3), 300. https://doi.org/10.21649/akemu.v18i3.420

Issue

Section

Articles
Loading...